Original Article A Phase I/II Trial to Evaluate the Technical Feasibility of Partial Breast Irradiation with Three-Dimensional Conformal Radiation Therapy in Korean Women with Stage I Breast Carcinoma: An Initial Report of the Korean Radiation Therapy Oncology Group (KROG) Study 0804
Jae-Uk Jeong, MD1, Jung Han Yoon, MD, PhD2, Min Ho Park, MD, PhD2, Mee Sun Yoon, MD, PhD1, Ju-Young Song, PhD1, Taek-Keun Nam, MD, PhD1, Woong-Ki Chung, MD, PhD1, Yong-Hyub Kim, MD1, Chang-Ok Suh, MD, PhD3, Sung-Ja Ahn, MD, PhD1,
Cancer Research and Treatment : Official Journal of Korean Cancer Association 2015;47(1):18-25.
DOI: https://doi.org/10.4143/crt.2013.202
Published online: August 21, 2014

1Department of Radiation Oncology, Chonnam National University Medical School, Gwangju, Korea

2Department of Surgery, Chonnam National University Medical School, Gwangju, Korea

3Department of Radiation Oncology, Yonsei University College of Medicine, Seoul, Korea

Correspondence: Sung-Ja Ahn, MD, PhD  Department of Radiation Oncology, Chonnam National University Hwasun Hospital, Chonnam National University Medical School, 322 Seoyang-ro, Hwasun 519-763, Korea 
Tel: 82-61-379-7200 Fax: 82-61-379-7249 E-mail: ahnsja@chonnam.ac.kr
• Received: October 18, 2013   • Accepted: December 17, 2013

Copyright © 2015 by the Korean Cancer Association

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted noncommercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

  • 12,514 Views
  • 79 Download
  • 4 Web of Science
  • 4 Crossref
  • 4 Scopus
prev next
  • Purpose
    This prospective study was designed to verify the technical feasibility of partial breast irradiation in breast cancer patients with small breasts, which are commonly encountered in Korean women.
  • Materials and Methods
    A total of 40 Gy, administered in 10 fractions on consecutive days (one fraction per day), was prescribed to the isocenters of the fields using three-dimensional conformal radiotherapy (3-DCRT). For all patients, treatment planning and dose parameters strictly adhered to the constraints set forth in the Radiation Therapy Oncology Group (RTOG) 0319 protocol. This study was designed such that if fewer than five of the first 42 evaluable patients received unacceptable scores, the treatment would be considered reproducible.
  • Results
    Ten treatment plans (23.8%) were determined to have major variations. There was no major variation in planning target volume (PTV) coverage. The ipsilateral and contralateral breast dose limitations were not met in four (9.5%) and four cases (9.5%), respectively. Major variations in ipsilateral and contralateral lung dose limitations were observed in two cases (4.8%). Major variations in the heart and thyroid dose limitations were observed in one (2.4%) and one case (2.4%), respectively. In multivariate analysis, a ratio of PTV to ipsilateral breast volume (PTV/IB) > 0.16 was the only significant factor that statistically affected major variations.
  • Conclusion
    We concluded that partial breast irradiation using 3-DCRT could not be reproduced in Korean breast cancer patients, particularly small-volumed breast surrogated as PTV/IB > 0.16. The dominant cause was the major variation in surrounding normal breast tissues.
A systematic review and meta-analysis of the current randomized trials to evaluate the role of accelerated partial breast irradiation (APBI) in eligible breast cancer patients suggested that APBI did not appear detrimental to survival and could be used as an alternative to whole breast irradiation (WBI) [1-3]. In selected patients, APBI, with fewer larger fractions delivered over a shorter time, could be an alternative to WBI because most relapses occur near the lumpectomy site [4-6].
The high “elsewhere” failure throughout the review of APBI frontier studies indicated the importance of appropriate patient selection and optimal APBI techniques [7]. A phase II study of APBI using proton therapy conducted in Korea showed excellent disease control [8]. However, only one center in Korea is equipped with proton therapy. Therefore, we adopted the three-dimensional conformal external beam radiotherapy (3-DCRT) technique for partial breast irradiation (PBI) because it is noninvasive, compared with brachytherapy or intraoperative therapy, and readily available in many institutions. The Radiation Therapy Oncology Group (RTOG) 0319 study reported that the technical feasibility of 3-DCRT for APBI was acceptable [9,10]. However, the 3-DCRT approach treats comparatively large normal breast tissue volumes while delivering relatively homogeneous doses to the target volume, as the basic underlying principle for the successful application of PBI is balanced radiation delivery between the planning target volume (PTV) and normal breast tissue.
We need to determine whether the technical feasibility confirmed in Western breast cancer patients could be equally applicable to Korean women, who have relatively small breasts compared to Western women. Therefore, we designed phase I and II clinical trials to evaluate the technical feasibility of 3-DCRT in Korean breast cancer patients, using the technique confirmed by RTOG 0319, to determine the types of variations that occur during PBI with 3-DCRT in small-volumed breast cancer patients. This is the first report of the technical feasibility of PBI with 3-DCRT in Korea.
1. Patient eligibility
We defined the inclusion criteria more strictly than that of RTOG 0319; the criteria were postmenopausal women, a unifocal tumor, tumor size ≤ 2 cm with a resection margin (RM) > 2 mm, negative sentinel or axillary nodes, hormone receptor positivity, and a good to excellent postoperative cosmesis score [11]. Patients with histological evidence of lymphovascular invasion, an extensive intraductal component, lobular histological type, ductal carcinoma in situ or lobular carcinoma in situ, and wound issues were excluded. Patients who received chemotherapy were also excluded. Radiation must have been initiated within eight weeks after surgery. The institutional review board approved this phase I/II trial, and all patients were prospectively enrolled and submitted consent forms.
2. Treatment techniques and imaging
Patients were immobilized in the supine position with both arms elevated upward, using a vacuum immobilization device. The ipsilateral whole breast and breast surgical scar were marked with a radiopaque rubber catheter. A treatment planning computed tomography (CT) scan (Hi-Speed FX/I, GE Medical Systems, Milwaukee, WI) was performed at a 3-mm scan width, and a contrast-enhanced image was obtained. The scan began at or above the mandible to beyond the breast tissue, including the entire lung. We used the Varian Eclipse treatment planning system (Varian Medical Systems, Palo Alto, CA). The following structures were contoured manually by the radiation oncologist: the lumpectomy cavity (LC) volume, PTV, ipsilateral breast, contralateral breast, thyroid, ipsilateral and contralateral lungs, and heart. We defined the target volume from the LC. To reduce the variability in LC contouring, radiation oncologists who participated in this study were educated and accustomed to the target volume definition guidelines. We assigned a clarity visualization score (CVS) to each patient, as follows; CVS-1, no cavity visualized; CVS-2, cavity visualized but significant uncertain margins; CVS-3, cavity visualized with some uncertain margins; CVS-4, cavity with mild heterogeneity on CT and mostly distinct margins; CVS-5, homogenous cavity with clearly identified margins [12]. The PTV was nonuniformly expanded 5-15 mm around the LC, depending on the safety margin widths along the six directions (superior, inferior, medial, lateral, superficial, and deep) described in the surgical pathologic report. We set a 1.5 cm expansion if the RM was < 1 cm, a 1-cm expansion if the RM was between ≥ 1 cm and < 2 cm, and a 0.5 cm expansion if the RM was ≥ 2 cm in all six directions. The PTV excluded the skin, was confined to a maximum of 5 mm under the skin, and vexcluded the underlying lung by lung-chest wall interface. Ipsilateral and contralateral breast included all tissue volume, excluding pectoralis muscles, chest wall muscles, and ribs as recommended by RTOG.
The treatment was administered with three-dimensional conformal fields, and intensity-modulated radiation therapy was not allowed. Three or four non-coplanar beams that used 6-MV photons or a mixed-modality plan that used a photon field with an en face electron field were used. The maximum dose should not exceed 110% of the prescribed dose. We followed the radiotheraphy delivery technique and dosimetric guidelines for dose volume constraints set forth by the RTOG 0319 study. A thorough review of the dose volume histogram of each patient was performed by both the radiation oncologist and physicist, and a quality assurance evaluation of the treatment plans was performed to determine the variations.
Each plan was scored as acceptable (per protocol), marginally acceptable (minor variation), or unacceptable (major variation). An acceptable plan showed 95% isodose surface covers 100% of the PTV, and all specified critical normal tissue dose limits have been met. A marginally acceptable plan showed 95% isodose surface covers between ≥ 95% to < 100% of the PTV, and all specified critical normal tissue dose limits fall within 5%. An unacceptable plan showed 95% isodose surface covers < 95% of the PTV, or any critical normal tissue dose limit exceeding 5% of the specified value. Each dose limitation for normal tissues followed the criteria described in the RTOG 0319 protocol [9].
In brief, less than 50% of the ipsilateral whole breast should receive ≥ 50% of the prescribed dose and less than 25% of the whole breast should receive the prescribed dose. The contralateral breast should receive less than 3% of the prescribed dose to any point. At most, less than 10% of the ipsilateral lung can receive 30% of the prescribed dose and less than 10% of the contralateral lung should receive 5% of the prescribed dose. In right-sided lesions, less than 5% of the heart should receive 5% of the prescribed dose in a while the volume of the heart receiving 5% of the prescribed dose should be less than that for treatment using conventional whole breast radiation with tangential fields in left-sided lesions. Maximum point dose in thyroid should be less than 3% of the prescribed dose.
A total of 40 Gy, given in 10 fractions on consecutive days, one fraction per day, were prescribed to the isocenters of the fields. Electronic portal imaging was monitored for the first session and repeated at the sixth radiotherapy session. No patients received chemotherapy, and adjuvant endocrine therapy was combined with radiotherapy in all patients.
3. Statistical considerations
Technical feasibility was the primary end point in this study. A final review of the plan as acceptable or marginally acceptable indicated that the technique is reproducible. An optimal two-stage Simon design was used for this trial. Let p-value be the true probability that the final review is acceptable or marginally acceptable. A p-value close to 1 implies that the radiation therapy is reproducible in a multi-center setting. If p-value is ≤ 80%, the goal is to have at most a 5% probability of concluding that the technique is reproducible. On the other hand, if p-value is ≥ 95%, the desired level, the goal is to have at most a 10% probability (five cases) of concluding that the technique is not reproducible in the first 42 evaluable treatment plans. Associations with pretreatment factors, including the tumor location, tumor side, ipsilateral breast volume, PTV volume, and PTV volume to ipsilateral breast volume ratio (PTV/IB) were tested using the chi-square test and logistic regression. Fisher’s exact test was used with small sample sizes (< 5 in any cells) in univariate analysis. All statistical data were processed using SPSS ver. 19.0 (SPSS Inc., Chicago, IL), and null hypotheses were rejected at p < 0.05.
1. Patient characteristics
A total of 42 patients were enrolled in this study between January 2009 and July 2013. The pretreatment characteristics are shown in Table 1. The median age of patients was 62 years, and the median tumor size was 11 mm. The tumor location was the outer or central quadrant for 37 patients (88%) and the inner quadrant for five patients (12%). The median LC volume at the time of treatment planning was 34.88 mL. The median PTV volume was 108.94 mL. The median ipsilateral whole breast volume was 784.18 mL. The median PTV/IB ratio was 0.1481 (Table 1).
2. Technical feasibility
Results of the plan evaluation with regard to dosimetric variations for the 42 patients, upon which the decision of feasibility was based, are shown in Table 2. There were no variations in 18 patients (42.9%) and minor variations in 14 patients (33.3%). Ten treatment plans (23.8%) were determined to have major variations, all of which were related to normal tissue dose constraints that exceeded 5% of the specified limit. There were no major variations in the PTV coverage. Figs. 1 and 2 show dose distribution and dose volume histogram of a case determined to have no major variation.
We analyzed the relationship between the tumor or normal tissue parameters and major variations in the treatment plans (Table 3). Of 12 patients with PTV > 130 mL, six (50%) had major variations. However, only four of 30 patients (13.3%) with PTV ≤ 130 mL had major variations. Major variations were observed more frequently in patients with PTV/IB > 0.16 (seven of 15 patients, 46.7%) than in patients with PTV/IB ≤ 0.16 (three of 27 patients, 11.1%). In multivariate analysis, the PTV/IB showed statistical significance affecting the major variation.
3. Dosimetric features in cases with major variations
Dose limitations in the ipsilateral and contralateral breast were not met in four (9.5%) and four cases (9.5%), respectively. Case no. 13 had major variations in both ipsilateral and contralateral breast dose limitations (Table 4). In this case, PTV was more than 130 mL (148.16 mL) and PTV/IB was more than 0.16 (0.19). In the four cases with major ipsilateral breast variations, all had PTV/IB > 0.16, and three cases had PTV > 130 mL. Six cases with PTV > 130 mL (case nos. 9, 13, 18, 27, 28, and 42) and seven with PTV/IB > 0.16 (case nos. 13, 18, 23, 27, 28, 31, and 42) had major variations. Among these cases, four with PTV > 130 mL (66.7%) and five with PTV/IB ratios > 0.16 (71.4%) did not fit the ipsilateral or contralateral breast dose limitations.
In the analysis of ipsilateral lung dose constraints, two cases (4.8%) showed major variations. Ipsilateral breast volumes < 900 mL and PTV/IB ratios > 0.16 were observed in both cases. In two cases with major variations in the contralateral lung dose limitation, we did not observe any specific features. Only one case (2.4%) had a major variation in the heart. In this case, the tumor was located in the lower central portion of the right breast. Although an electron beam was used in part in this case, more than 2 Gy were delivered to the right atrium. A thyroid dose limitation was observed in one case (2.4%). In this case, the LC was in the upper quadrant of the left breast, near the thyroid gland (Table 4).
As clinical experiences with PBI increase, it is certain that patient subsets might be adequately treated with PBI, with acceptably low recurrence rates. As PBI currently remains an investigational approach and is not yet a standard therapy, we designed this study to include a more rigorously selected group of patients than would be allowed by the “suitable” criteria suggested in the American Society for Radiation Oncology consensus statement [2]. We further limited eligible patients for this study to those without wound issues and chemotherapy and those with good to excellent cosmesis scores at the time of radiotherapy.
Delineation of the optimal treatment volume required to obtain equal or better local control rates, compared with WBI, after conservative surgery is crucial to the precision of conformal radiation therapy planning and the validity of clinical trial results [13]. The issue of surgical clips during LC visualization when performing PBI with 3-DCRT has been highlighted as mandatory [14]. However, surgeons at our institute do not use surgical clips; therefore, we used the LC to define the PBI treatment target. In a study of change of the serial morphologic cavity after surgery, the optimal time to delineate the target volume from the excision cavity in the planning CT scan for PBI was reported to be within eight weeks [15]. Based on this data, we limited study entry to patients who could undergo PBI within eight weeks after surgery. Variability among radiation oncologists regarding delineation of the postsurgical LC could be decreased with a CVS ≥ 3 [12]. In our study, CVS < 3 was observed in only one patient. For this patient, when defining the LC, we obtained a consensus among radiation oncologists to include all surgical tissue extensions. We chose a different method to define the PTV than that used in the RTOG 0319 study. We differentiated the optimal margin to be included in the PTV according to the surgical safety margin. This was meant to provide the same final margin to the gross tumor area in all patients, irrespective of the extent of volume surrounding the gross tumor that was surgically removed.
The widespread acceptance of APBI using the 3-DCRT approach is likely because it is totally noninvasive and delivers a homogenous dose distribution to the target volume. In our study, PTV was defined to exclude the skin (5 mm beneath the skin surface) in order to reduce skin toxicity [16] and there was no case in which skin should be included in PTV due to positive or close superficial RM. And, there was no major variation in the PTV coverage because the beam-aperture was composed to cover the PTV. The majority of the major variation was caused by the inability to fit the dose constraints on normal breast tissue. The surrounding normal tissue dose-volume relationship is a critical underlying factor, in accordance with the dosimetric guidelines. Regarding the PTV/IB ratio, large target volumes (PTV) contribute to an increased risk of severe fibrosis and poor cosmesis [17]. Leonard et al. [18] reported a clear and statistically significant association between the dose-volume parameters and late moderate-to-severe breast toxicity. The relative volumes of breast tissue that receive low and high percentages (V5%-V100%) of the prescribed dose were associated with the risk of subcutaneous fibrosis, and the volume receiving 50%, 80%, and 100% (V50%-V100%) of the prescribed dose was associated with fair/poor cosmesis scores. These results generally demonstrate a higher incidence of late normal tissue toxicity compared with those reported in the RTOG trials with 3-DCRT APBI [9,10]. In order to substantially reduce the risk of adverse normal tissue effects, the authors suggested the use of more restrictive dose-volume constraints than those currently defined by the National Surgical Adjuvant Breast and Bowel Project (NSABP) B39/RTOG 0419 trial. However, the authors noted a potential limitation of the retrospective analysis in that a baseline postsurgical evaluation was not performed, thus making it difficult to assess the true adverse effects of radiation-induced normal tissue toxicity; therefore, additional results are needed before these suggestions can be accepted.
In our study, the ipsilateral breast dose limitations of V50% < 50% were not acceptable in four cases (9.5%), while, there was no violation of the ipsilateral breast dose limitation in the RTOG 0319 study [9]. As a possible explanation for the discordance between the two studies, we can consider the differences in breast size. In a recent phase I/II study of APBI with 3-DCRT in Western countries [19], the median ipsilateral breast volume, 1,698 mL (range, 647 to 3,627 mL), was 2-fold larger than that in our study, 784 mL (range, 337 to 1,446 mL). The median V50% of the ipsilateral breast was 41.9% in our study (data was not shown), and this parameter was larger than that reported in Western countries (37.9%). In a Korean study of dosimetric comparison of PBI with four different techniques, mean V50% in 3-DCRT was 57.6%, and this parameter is similar to our data [20], meaning that relatively larger portions of the normal breast tissue might be exposed to low-dose irradiation (V50%) in Korean women with small breasts. In addition, four cases (9.5%) showed major variations in contralateral breast dose limitations. In these cases, only one had an ipsilateral breast dose violation. There was a limitation in the gantry angle of the linear accelerator with regard to fitting the ipsilateral breast dose limitation, and this could cause partial contralateral breast irradiation in small breasts. Comparatively large normal breast tissue could be exposed to low-dose irradiation by use of PBI with 3-DCRT while relatively homogeneous doses could be delivered to the target volume. Therefore, the ipsilateral breast dose limitation in the RTOG 0319 study might not be applicable to Korean women with smallvolumed breasts.
Although we could not draw the same conclusion regarding technical feasibility as in the RTOG 0319 study, we can conclude that the PTV/IB ratio affects the technical feasibility of 3-DCRT planning for PBI and it can be performed in Korean breast cancer patients with PTV/IB ≤ 0.16, even those with small breasts. In addition, further study is needed in order to determine feasible dose limitations, particularly breast limitation, when PBI with external-beam irradiation is administered to small-breasted patients.

Conflict of interest relevant to this article was not reported.

Fig. 1.
Dose distribution of partial breast irradiation using two coplanar and two non-coplanar photon beams.
crt-2013-202f1.gif
Fig. 2.
Dose volume histogram of partial breast irradiation using two coplanar and two non-coplanar photon beams. Lumpectomy cavity (purple), planning target volume (red), ipsilateral breast (yellow), contralateral breast (cyan), ipsilateral lung (blue), contralateral lung (green), and heart (orange) are depicted.
crt-2013-202f2.gif
Table 1.
Patient characteristics
PBI (n=42)
Age (yr)
 Median 62.0
 Range 50-76
pT stage
 mi 2 (4.8)
 1a 2 (4.8)
 1b 14 (33.3)
 1c 24 (57.1)
Tumor size (cm)
 Median 1.1
 Range 0.1-1.9
Tumor side
 Right 12 (28.6)
 Left 30 (71.4)
Tumor location
 Upper outer 26 (61.8)
 Upper inner 3 (7.1)
 Lower outer 4 (9.5)
 Lower inner 2 (4.9)
 Upper central 6 (14.3)
 Lower central 1 (2.4)
Cavity visualization score
 2 1 (2.4)
 3 11 (26.2)
 4 19 (45.2)
 5 11 (26.2)
Lumpectomy cavity (mL)
 Median 34.88
 Range 15.40-86.73
PTV (mL)
 Median 108.94
 Range 59.59-185.67
Ipsilateral breast volume (mL)
 Median 784.18
 Range 337.58-1,476.60
PTV/IB
 Median 0.1481
 Range 0.07-0.23

Values are presented as number (%). PBI, partial breast irradiation; pT stage, pathologic T stage; PTV, planning target volume; PTV/IB, planning target volume to ipsilateral breast volume ratio.

Table 2.
Technical feasibility of radiation treatment planning using three-dimensional conformal radiotherapy
PTV coverage Ipsilateral breast Contralateral breast Ipsilateral lung Contralateral lung Heart Thyroid Overall
No variation 26 (61.9) 36 (85.7) 33 (78.6) 37 (88.1) 40 (95.2) 41 (97.6) 41 (97.6) 18 (42.9)
Minor variation 16 (38.1) 2 (4.8) 5 (11.9) 3 (7.1) 0 0 0 14 (33.3)
Major variation 0 4 (9.5) 4 (9.5) 2 (4.8) 2 (4.8) 1 (2.4) 1 (2.4) 10 (23.8)

Values are presented as number (%). PTV, planning target volume.

Table 3.
Pretreatment factors impacting technical feasibility using three-dimensional conformal radiotherapy
p-value
Univariate analysisa) Multivariate analysisb)
Tumor location (lower) 0.203 -
Tumor location (inner) 0.341 -
Tumor side (left) 0.600 -
Ipsilateral breast volume (> 900 mL) 0.476 -
PTV (> 130 mL) 0.020 -
PTV/IB (> 0.16) 0.014 0.015 (OR, 7.0; 95% CI, 1.4-33.7)

Values are presented as number (%). PTV, planning target volume; PTV/IB, planning target volume to ipsilateral breast volume ratio; OR, odds ratio; CI, confidence interval.

a) Fisher’s exact test was used due to small sample size (less than five cases in any cell),

b) Logistic regression test was performed.

Table 4.
Dosimetric features in cases with major variations of treatment planning using three-dimensional conformal radiotherapy
Case no. PTV (mL) IB (mL) PTV/IB Side Location Location (inner) Major variation
2 120.82 818.22 0.147 Left Lower No Contralateral breast
9 141.47 1,446.69 0.097 Left Upper No Thyroid
13 148.16 763.54 0.194 Left Upper No Ipsilateral breast, contralateral breast
18 155.32 864.37 0.179 Right Upper No Ipsilateral breast
23 63.59 337.58 0.188 Right Upper No Contralateral breast
27 134.42 807.71 0.166 Right Lower No Ipsilateral lung, contralateral lung, heart
28 175.94 764.87 0.23 Left Lower Yes Ipsilateral breast
31 88.28 528.81 0.166 Left Upper Yes Ipsilateral breast, ipsilateral lung
36 74.08 487.98 0.151 Left Upper No Contralateral lung
42 167.72 1,031.22 0.162 Left Upper No Contralateral breast

PTV, planning target volume; IB, ipsilateral breast volume; PTV/IB, planning target volume to ipsilateral breast volume ratio.

  • 1. Polgar C, Fodor J, Major T, Nemeth G, Lovey K, Orosz Z, et al. Breast-conserving treatment with partial or whole breast irradiation for low-risk invasive breast carcinoma: 5-year results of a randomized trial. Int J Radiat Oncol Biol Phys. 2007;69:694–702. ArticlePubMed
  • 2. Smith BD, Arthur DW, Buchholz TA, Haffty BG, Hahn CA, Hardenbergh PH, et al. Accelerated partial breast irradiation consensus statement from the American Society for Radiation Oncology (ASTRO). Int J Radiat Oncol Biol Phys. 2019;74:987–1001. Article
  • 3. Valachis A, Mauri D, Polyzos NP, Mavroudis D, Georgoulias V, Casazza G. Partial breast irradiation or whole breast radiotherapy for early breast cancer: a meta-analysis of randomized controlled trials. Breast J. 2010;16:245–51. ArticlePubMed
  • 4. Fowble B, Solin LJ, Schultz DJ, Rubenstein J, Goodman RL. Breast recurrence following conservative surgery and radiation: patterns of failure, prognosis, and pathologic findings from mastectomy specimens with implications for treatment. Int J Radiat Oncol Biol Phys. 1990;19:833–42. ArticlePubMed
  • 5. Smith TE, Lee D, Turner BC, Carter D, Haffty BG. True recurrence vs. new primary ipsilateral breast tumor relapse: an analysis of clinical and pathologic differences and their implications in natural history, prognoses, and therapeutic management. Int J Radiat Oncol Biol Phys. 2000;48:1281–9. ArticlePubMed
  • 6. Morrow M. Rational local therapy for breast cancer. N Engl J Med. 2002;347:1270–1. ArticlePubMed
  • 7. Ribeiro GG, Magee B, Swindell R, Harris M, Banerjee SS. The Christie Hospital breast conservation trial: an update at 8 years from inception. Clin Oncol (R Coll Radiol). 1993;5:278–83. ArticlePubMed
  • 8. Chang JH, Lee NK, Kim JY, Kim YJ, Moon SH, Kim TH, et al. Phase II trial of proton beam accelerated partial breast irradiation in breast cancer. Radiother Oncol. 2013;108:209–14. ArticlePubMed
  • 9. Vicini F, Winter K, Straube W, Wong J, Pass H, Rabinovitch R, et al. A phase I/II trial to evaluate three-dimensional confor-mal radiation therapy confined to the region of the lumpectomy cavity for Stage I/II breast carcinoma: initial report of feasibility and reproducibility of Radiation Therapy Oncology Group (RTOG) Study 0319. Int J Radiat Oncol Biol Phys. 2005;63:1531–7. ArticlePubMed
  • 10. Vicini F, Winter K, Wong J, Pass H, Rabinovitch R, Chafe S, et al. Initial efficacy results of RTOG 0319: three-dimensional conformal radiation therapy (3D-CRT) confined to the region of the lumpectomy cavity for stage I/II breast carcinoma. Int J Radiat Oncol Biol Phys. 2010;77:1120–7. ArticlePubMedPMC
  • 11. Harris JR, Levene MB, Svensson G, Hellman S. Analysis of cosmetic results following primary radiation therapy for stages I and II carcinoma of the breast. Int J Radiat Oncol Biol Phys. 1979;5:257–61. ArticlePubMed
  • 12. Landis DM, Luo W, Song J, Bellon JR, Punglia RS, Wong JS, et al. Variability among breast radiation oncologists in delineation of the postsurgical lumpectomy cavity. Int J Radiat Oncol Biol Phys. 2007;67:1299–308. ArticlePubMed
  • 13. Smitt MC, Nowels KW, Zdeblick MJ, Jeffrey S, Carlson RW, Stockdale FE, et al. The importance of the lumpectomy surgical margin status in long-term results of breast conservation. Cancer. 1995;76:259–67. ArticlePubMed
  • 14. Dzhugashvili M, Pichenot C, Dunant A, Balleyguier C, Delaloge S, Mathieu MC, et al. Surgical clips assist in the visualization of the lumpectomy cavity in three-dimensional conformal accelerated partial-breast irradiation. Int J Radiat Oncol Biol Phys. 2010;76:1320–4. ArticlePubMed
  • 15. Petersen RP, Truong PT, Kader HA, Berthelet E, Lee JC, Hilts ML, et al. Target volume delineation for partial breast radiotherapy planning: clinical characteristics associated with low interobserver concordance. Int J Radiat Oncol Biol Phys. 2007;69:41–8. ArticlePubMed
  • 16. Vicini FA, Beitsch PD, Quiet CA, Keleher A, Garcia D, Snider HC, et al. First analysis of patient demographics, technical reproducibility, cosmesis, and early toxicity: results of the American Society of Breast Surgeons MammoSite breast brachytherapy trial. Cancer. 2005;104:1138–48. ArticlePubMed
  • 17. Hepel JT, Tokita M, MacAusland SG, Evans SB, Hiatt JR, Price LL, et al. Toxicity of three-dimensional conformal radiotherapy for accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2009;75:1290–6. ArticlePubMed
  • 18. Leonard KL, Hepel JT, Hiatt JR, Dipetrillo TA, Price LL, Wazer DE. The effect of dose-volume parameters and interfraction interval on cosmetic outcome and toxicity after 3-dimensional conformal accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2013;85:623–9. ArticlePubMed
  • 19. Bergom C, Prior P, Kainz K, Morrow NV, Ahunbay EE, Walker A, et al. A phase I/II study piloting accelerated partial breast irradiation using CT-guided intensity modulated radiation therapy in the prone position. Radiother Oncol. 2013;108:215–9. ArticlePubMed
  • 20. Moon SH, Shin KH, Kim TH, Yoon M, Park S, Lee DH, et al. Dosimetric comparison of four different external beam partial breast irradiation techniques: three-dimensional conformal radiotherapy, intensity-modulated radiotherapy, helical tomotherapy, and proton beam therapy. Radiother Oncol. 2009;90:66–73. ArticlePubMed

Figure & Data

REFERENCES

    Citations

    Citations to this article as recorded by  
    • Assessment of Eligibility and Utilization of Accelerated Partial Breast Irradiation in Korean Breast Cancer Patients (KROG 22-15)
      Seok-Joo Chun, Ji Hwan Jo, Yong Bae Kim, Sangjoon Park, Sung-Ja Ahn, Su Ssan Kim, Kyubo Kim, Kyung Hwan Shin
      Cancer Research and Treatment.2024; 56(2): 549.     CrossRef
    • Intensity Modulated Radiotherapy and Volumetric Modulated Arc Therapy in the Treatment of Breast Cancer: An Updated Review
      Jee Suk Chang, Ji Hyun Chang, Nalee Kim, Yong Bae Kim, Kyung Hwan Shin, Kyubo Kim
      Journal of Breast Cancer.2022; 25(5): 349.     CrossRef
    • Who are the optimal candidates for partial breast irradiation?
      Ji Hyeon Joo, Yongkan Ki, Hosang Jeon, Dong Woon Kim, Jinhong Jung, Su Ssan Kim
      Asia-Pacific Journal of Clinical Oncology.2021; 17(4): 305.     CrossRef
    • First Experience in Korea of Stereotactic Partial Breast Irradiation for Low-Risk Early-Stage Breast Cancer
      Won Hee Lee, Jee Suk Chang, Min Jung Kim, Vivian Youngjean Park, Jung Hyun Yoon, Se Young Kim, Jee Ye Kim, Hyung Seok Park, Seung Il Kim, Young Up Cho, Byeong Woo Park, Yong Bae Kim
      Frontiers in Oncology.2020;[Epub]     CrossRef

    • PubReader PubReader
    • ePub LinkePub Link
    • Cite
      CITE
      export Copy Download
      Close
      Download Citation
      Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

      Format:
      • RIS — For EndNote, ProCite, RefWorks, and most other reference management software
      • BibTeX — For JabRef, BibDesk, and other BibTeX-specific software
      Include:
      • Citation for the content below
      A Phase I/II Trial to Evaluate the Technical Feasibility of Partial Breast Irradiation with Three-Dimensional Conformal Radiation Therapy in Korean Women with Stage I Breast Carcinoma: An Initial Report of the Korean Radiation Therapy Oncology Group (KROG) Study 0804
      Cancer Res Treat. 2015;47(1):18-25.   Published online August 21, 2014
      Close
    • XML DownloadXML Download
    Figure
    • 0
    • 1
    A Phase I/II Trial to Evaluate the Technical Feasibility of Partial Breast Irradiation with Three-Dimensional Conformal Radiation Therapy in Korean Women with Stage I Breast Carcinoma: An Initial Report of the Korean Radiation Therapy Oncology Group (KROG) Study 0804
    Image Image
    Fig. 1. Dose distribution of partial breast irradiation using two coplanar and two non-coplanar photon beams.
    Fig. 2. Dose volume histogram of partial breast irradiation using two coplanar and two non-coplanar photon beams. Lumpectomy cavity (purple), planning target volume (red), ipsilateral breast (yellow), contralateral breast (cyan), ipsilateral lung (blue), contralateral lung (green), and heart (orange) are depicted.
    A Phase I/II Trial to Evaluate the Technical Feasibility of Partial Breast Irradiation with Three-Dimensional Conformal Radiation Therapy in Korean Women with Stage I Breast Carcinoma: An Initial Report of the Korean Radiation Therapy Oncology Group (KROG) Study 0804
    PBI (n=42)
    Age (yr)
     Median 62.0
     Range 50-76
    pT stage
     mi 2 (4.8)
     1a 2 (4.8)
     1b 14 (33.3)
     1c 24 (57.1)
    Tumor size (cm)
     Median 1.1
     Range 0.1-1.9
    Tumor side
     Right 12 (28.6)
     Left 30 (71.4)
    Tumor location
     Upper outer 26 (61.8)
     Upper inner 3 (7.1)
     Lower outer 4 (9.5)
     Lower inner 2 (4.9)
     Upper central 6 (14.3)
     Lower central 1 (2.4)
    Cavity visualization score
     2 1 (2.4)
     3 11 (26.2)
     4 19 (45.2)
     5 11 (26.2)
    Lumpectomy cavity (mL)
     Median 34.88
     Range 15.40-86.73
    PTV (mL)
     Median 108.94
     Range 59.59-185.67
    Ipsilateral breast volume (mL)
     Median 784.18
     Range 337.58-1,476.60
    PTV/IB
     Median 0.1481
     Range 0.07-0.23
    PTV coverage Ipsilateral breast Contralateral breast Ipsilateral lung Contralateral lung Heart Thyroid Overall
    No variation 26 (61.9) 36 (85.7) 33 (78.6) 37 (88.1) 40 (95.2) 41 (97.6) 41 (97.6) 18 (42.9)
    Minor variation 16 (38.1) 2 (4.8) 5 (11.9) 3 (7.1) 0 0 0 14 (33.3)
    Major variation 0 4 (9.5) 4 (9.5) 2 (4.8) 2 (4.8) 1 (2.4) 1 (2.4) 10 (23.8)
    p-value
    Univariate analysisa) Multivariate analysisb)
    Tumor location (lower) 0.203 -
    Tumor location (inner) 0.341 -
    Tumor side (left) 0.600 -
    Ipsilateral breast volume (> 900 mL) 0.476 -
    PTV (> 130 mL) 0.020 -
    PTV/IB (> 0.16) 0.014 0.015 (OR, 7.0; 95% CI, 1.4-33.7)
    Case no. PTV (mL) IB (mL) PTV/IB Side Location Location (inner) Major variation
    2 120.82 818.22 0.147 Left Lower No Contralateral breast
    9 141.47 1,446.69 0.097 Left Upper No Thyroid
    13 148.16 763.54 0.194 Left Upper No Ipsilateral breast, contralateral breast
    18 155.32 864.37 0.179 Right Upper No Ipsilateral breast
    23 63.59 337.58 0.188 Right Upper No Contralateral breast
    27 134.42 807.71 0.166 Right Lower No Ipsilateral lung, contralateral lung, heart
    28 175.94 764.87 0.23 Left Lower Yes Ipsilateral breast
    31 88.28 528.81 0.166 Left Upper Yes Ipsilateral breast, ipsilateral lung
    36 74.08 487.98 0.151 Left Upper No Contralateral lung
    42 167.72 1,031.22 0.162 Left Upper No Contralateral breast
    Table 1. Patient characteristics

    Values are presented as number (%). PBI, partial breast irradiation; pT stage, pathologic T stage; PTV, planning target volume; PTV/IB, planning target volume to ipsilateral breast volume ratio.

    Table 2. Technical feasibility of radiation treatment planning using three-dimensional conformal radiotherapy

    Values are presented as number (%). PTV, planning target volume.

    Table 3. Pretreatment factors impacting technical feasibility using three-dimensional conformal radiotherapy

    Values are presented as number (%). PTV, planning target volume; PTV/IB, planning target volume to ipsilateral breast volume ratio; OR, odds ratio; CI, confidence interval.

    Fisher’s exact test was used due to small sample size (less than five cases in any cell),

    Logistic regression test was performed.

    Table 4. Dosimetric features in cases with major variations of treatment planning using three-dimensional conformal radiotherapy

    PTV, planning target volume; IB, ipsilateral breast volume; PTV/IB, planning target volume to ipsilateral breast volume ratio.


    Cancer Res Treat : Cancer Research and Treatment
    Close layer
    TOP